OPERATION ,CLEANING, CALIBRATION AND MAINTENANCE OF INFRARED MOISTURE ANALYSER
OPERATION , CLEANING, CALIBRATION, AND MAINTENANCE OF INFRARED MOISTURE ANALYSER
1) PURPOSE : To describe the method of Operation, Cleaning, Calibration and Maintenance of an Infrared (IR) Moisture Analyser
2) SCOPE : This SOP is applicable for the Operation, Cleaning, Calibration and Maintenance of IR moisture analyser Make Ohaus, Model-MB23,Capacity 110 g ,Accuracy 0.001% having Temperature Range 500C to 1600C installed in the QC department
3) FREQUENCY :
- Operation : As and when required
- Cleaning : Daily before & after operation
- Calibration : Daily for balance & yearly for temperature sensor
- Maintenance : As and when required, mainly during break down
4) RESPONSIBILITY : Executive QC, Officer QC
5) EQUIPMENT /MATERIALS REQUIRED : IR Moisture Balance, Clean spatula, Certified Calibration weights, Clean Mortar & Pestle, Clean & Dry Brush.
6) PRECAUTIONS :
- Avoid the unnecessary burning of IR lamp.
- Do not handle the moisture analyser roughly.
- Adjust the spirit levels by using levelling screws of the balance.
- Use voltage stabiliser.
- Do not weigh the magnetized samples.
7) PROCEDURE :
- Operation :
- Switch “ON” the mains.
- Check the spirit level of IR Moisture Analyser.
- Press the POWER/BRK’ key to display grams, now the machine is in the measurement mode and is ready to measure the sample.
- Press “TARE” and wait till 0.000 appears on the LCD.
- Crush the sample in mortal for uniform granules & add the1.00 gm sample.
- Note the weight of the sample displays on the LCD.
- Spread the sample uniformly such that it has thickness about 1 mm.
- After completion of the drying the results appear on the LCD as percentage.
8) Cleaning :
- Wipe all the traces of organic solvent/water /moisture by tissue paper.
- Clean all the dust from the pan by tissue paper.
- Cover the instrument by PVC cover
- Calibration :
- Calibrate the IR moisture balance with certified weight once daily.
Sr. No. | Standard Weight in gm | Tolerance |
1. | 200 g | 199.800 – 200.200 g |
2. | 100 g | 99.900 – 100.100 g |
3. | 50 g | 49.950 – 50.050 g |
4. | 20 g | 19.980 – 20.020 g |
5. | 10 g | 9.990 – 10.010 g |
6. | 5 g | 4.995 – 5.005 g |
7. | 2 g | 1.998 – 2.002 g |
8. | 1 g | 0.999 – 1.001 g |
9. | 0.5 g | 0.499 – 0.501 g |
10. | 0.2 g | 0.199 – 0.201 g |
11. | 0.1 g | 0.099 – 0.101 g |
12. | 0.05 g | 0.049 – 0.051 g |
13. | 0.02 g | 0.019 – 0.021 g |
- Calibrate the temperature sensor by outside authorised agency once in a year.
- Moisture value of sample (Di- Sodium Tartrate)
Sr no. | Sample weight (Disodium tartrate | Temp ⁰C | Time in min | Moisture Value(%M) | Acceptance criteria 15.0 to 16.0 % w/w | Remarks satisfactory/unsatisfactory |
1. | 1.0 g | 150 ⁰C | Auto mode | 15.0-16.0 % | ||
2. | 1.0 g | 150 ⁰C | Auto mode | 15.0-16.0 % | ||
3. | 1.0 g | 150 ⁰C | Auto mode | 15.0-16.0 % | ||
4. | 1.0 g | 150 ⁰C | Auto mode | 15.0-16.0 % | ||
5. | 1.0 g | 150 ⁰C | Auto mode | 15.0-16.0 % |
9)Maintenance :
If there is any breakdown, contact to the maintenance department for minor
If there are major faults, contact to authorised service engineers.
10) ENCLOSURE :
Calibration record of Infrared Moisture Analyser :
Daily logbook of Infrared Moisture Analyser :