CALIBRATION RECORD OF INFRARED MOISTURE BALANCE

Name of instrument – Department  –

Make no-

Date of calibration –            

Model no   –

Next due date –
Identification no –     Frequency –

 

CALIBRATION OF IR BALANCE

Calibrate the IR moisture balance with certified weight once in a daily.

Sr. No.Standard Weight in gmObserved Weight in gmTolerance  + 0.5 %
11.0 0.995  to  1.005
22.0 1.99    to   2.01
35.0 4.975  to   5.025
410.0 9.950  to  10.05

REMARK The  observed value of standard weight  are  found / are not found within tolerance limit &  hence it is accepted. 

CONCLUSION :  The instrument is calibrated and performs satisfactorily/ instrument is out of order and corrective action is recommended.

 

CORRECTIVE ACTION: _________________________________________________________________________

 

Calibration done by : _____________                                     Checked by : _____________

Sign & date               : _____________                                   Sign & date : ______________